• Home > Journal Search > Past Issues

Past Issues

Endnote RefWorks Scholar's Aid Excel TXT
Vol.11, No.2, 235 ~ 244, 2008
Title
Development of continuous blood pressure measurement system using ECG and PPG
 
Abstract
This study is to develop automatic extraction system of continuous blood pressure using ECG (Electrocardiogram) and PPG(Photoplethysmography) for u-health care technology. PTT (Pulse Transit Time) was determined from peak difference between ECG and PPG and its inverse made to get blood pressure. Since the peaks were vulnerable to be contaminated from noise and variation of amplitude, this study developed the adaptive algorithm for peak calculation in any noise condition. The developed method of the adaptive peak calculation was proven to make the standard deviations of PPT decrease to 28% and the detection of noise increase to 18%. Also, the correlation model such as blood pressure = -0.044 · PTT + 133.592 has successfully been determined for predicting the continuous pressure measured without using cuff but with using PPG and ECG, only.
Key Words
PTT, Pulse Transit Time, Blood Pressure, Photoplethsmography, Electrocardiogram, Adaptive algorithm